Page 16
5. mV/ORP Calibration
The mV/ORP section is factory calibrated. There is no user-calibration
facility for this mode.
6. Relative mV/ORP Calibration
Select Relative mV/ORP mode when measurements relative to a known
standard are required. Calibration of the Relative mV/ORP mode is simply
a matter of zeroing the reading when the sensor is in the known standard.
1. Plug the Redox sensor into the
Channel 1
or
Channel 2
socket.
Temperature compensation is not applied in Relative mV/ORP mode,
so the temperature sensor does not need to be connected.
2. Switch the meter on.
3. Select Relative mV/ORP Mode
Either (
→
→
→
→
F3:Mode
→
→
→
→
F1:Chan1
→
→
→
→
F3:Rel mV/ORP
).
or
(
→
→
→
→
F3:Mode
→
→
→
→
F2:Chan2
→
→
→
→
F3:Rel mV/ORP
).
4. The display should now be showing Relative mV/ORP on the top line
with absolute mV/ORP or date and time on the bottom line. For
example…
500.0R 7.00pH
660 F4:Zeros
↔
500.0R 7.00pH
31/12 12:00:00
Press
to alternate between absolute mV/ORP or the clock on the
bottom line.
5. Remove the wetting cap from the Redox sensor.
6. Rinse the sensor in distilled water and blot dry.
7. Place the Redox sensor into a sample of the known standard. Ensure
that the platinum tip and reference junction are both covered, as per
the diagram over the page.