Page 29
11.6 Data Format
Data is returned to the RS232 Port by the
WP-80D
in the following format.
A “
•
” shown anywhere in this section denotes one space.
LLLL
PPPPPPuuu
ppppppuuu
TTTTTTuuu
dd/mm/yy
hh:mm:ss
LLLL
is the Log Number. Maximum 4 characters, right
justified. The
WP-80D
sends a Zero for instant readings
(section 11.2)
PPPPPP
is Channel 1 pH/mV/ORP data. 6 characters, right
justified.
pppppp
is Channel 2 pH/mV/ORP data. 6 characters, right
justified.
uuu
is the pH/mV/ORP unit description, which can be any of
the following…
pH•
for pH readout.
mV•
for Millivolts readout.
mVR
for Relative Millivolts readout.
TTTTTT
is Temperature data, 6 characters, right justified.
uuu
is the Temperature unit description, which can be either
of the following…
oC•
for real Temperature data.
oCm
for manual Temperature
compensation values.
dd/mm/yy
is the date, month and year data.
hh:mm:ss
is the hours, minutes and seconds data.
When requested by a PC with the ?D or ?R commands (section 11.5), the
data is terminated with a carriage return.
When the data is sent by the
WP-80D
using the Print function (section
9.4) or the Send function (section 11.2) the data ends with a carriage
return and a line feed.