Page 27
11. RS232 Port
11.1 Setting the Baud Rate
1. Select the RS232 Set-up menu (
→
→
→
→
F4:Setup
→
→
→
→
F3:Baud
).
2. The available baud rates are listed on the display.
F1:300 F2:1200
>F3:9600 F4:19200
The arrow shows the current selection.
3. Press
to select 300 baud
Press
to select 1200 baud
Press
to select 9600 baud
Press
to select 19200 baud.
Press
to quit and retain the current setting.
11.2 Sending Readings to the RS232 Port
Press
to instantly send readings to the RS232 port whenever the
WP-80D
is in normal run mode. This function is disabled if the automatic
logging period is set to greater than zero (see section 10).
Records can be sent directly to the RS232 port rather than stored in
memory during automatic data logging. See section 10 for details.
11.3 RS232 Configuration
The
WP-80D
RS232 configuration is 8 bits, No Parity, 1 Stop Bit,
XON/XOFF Protocol.
11.4 Communication and Statistical Software
Communication between the
WP-80D
and a PC can be handled with any
RS232 communication software. A TPS communication software package
for Windows
is optionally available (part number 130086).
Once the data is saved to disk, the next problem is how to use it. The
data sent by the
WP-80D
is formatted in columns that can be imported by
programs such as Microsoft
®
Excel
®
and Lotus 123
®
.
Help on importing the data into Microsoft
®
Excel
®
is provided in section
11.8 and also the “excel.txt” file in the folder where you installed the
WinTPS program.