Page 37
16. Initialising the WP-80D
If you need to phone or fax TPS for any further technical assistance, the
version number of your
WP-80D
firmware may of benefit to us. The
version number is displayed by the
WP-80D
at turn-on.
If the calibration settings of the
WP-80D
exceed the allowable limits, the
unit may need to be initialised to factory default values. This action may
be required if the electrode is replaced.
To initialise the
WP-80D
:
1. Switch the
WP-80D
off.
2. Press and hold the
key while switching the
WP-80D
back on.
3. The following messages should be displayed:
Initialized
MUST ReCalibrate
then :
WP80Ds V2.0 S1234
Dual
pH/mV/ORP,
Temp
(The “s” after
WP-80D
is shown when the RS232 serial port option is
fitted)
4. The meter then displays pH and Temperature. Note that the decimal
points have been replaced with a “
∗
”, to indicate that the unit requires
recalibration.
17. Instrument firmware version number.
If you need to phone or fax TPS for any further technical assistance, the
version number of your
WP-80D
firmware may of benefit to us. The
version number is displayed by the
WP-80D
at turn-on.