14. Initialising the 900-I3
If the calibration settings of the
900-I3
exceed the allowable limits, the unit may need to be
initialised to factory default values. This action may be required if the electrode is replaced.
To initialise the
900-I3
…
1.
Switch the
900-I3
off.
2.
Press and hold the
key while switching the
900-I3
back on.
3.
The following messages should be displayed…
Initializing
Initializing
then…
System Initialized
System Initialized
Unit should be Re-calibrated before use
Unit should be Re-calibrated before use
then…
900I3srV3.0 R1234 (C) 1998 TPS Pty Ltd
900I3srV3.0 R1234 (C) 1998 TPS Pty Ltd
Three Channel Ion/pH/mV and Temperature
Three Channel Ion/pH/mV and Temperature
The “
s
s
” after
900I3
is shown when the RS232 serial Port option is fitted.
The “
r
r
” after
900I3
is shown when the Recorder Port option is fitted.
4.
The meter then displays pH, Temperature, Date and Time. Note that the decimal points have
been replaced with a “
∗∗
”, to indicate that the unit requires re-calibration.
15. Instrument firmware version number
If you need to phone or fax TPS for any further technical assistance, the version number of your
900-I3
firmware may of benefit to us. The version number is displayed by the
900-I3
at turn-on.