38
4.
(a) If
F1:Ion
F1:Ion
was selected at step 3 above, the
900-I3
will now ask you to enter the Specific
Ion reading that will produce full scale on the Recorder Port. For example…
Enter Full Scale :
Enter Full Scale : 1
1000 ppM
000 ppM
For ppM, ppK and % readouts, type in the Specific Ion reading, including decimal point,
that will produce full scale.
For Exponential mode, type in 3 significant digits of the value that will produce full scale.
The decimal point is fixed for you in this mode.
Now press
for 10
+
or press
for 10
–
.
Enter the power. This can be entered as 1 or 2 digits (e.g. “
04
04
” is treated the same as “
4
4
”).
Use the
key to make any corrections.
Press
to save the value.
Alternatively, press
to quit without changing the current setting.
The
900-I3
will confirm the setting before returning to normal measurement mode. For
example…
Channel 1 Ion selected,
Channel 1 Ion selected,
1000.ppM = 2 volts
1000.ppM = 2 volts