24
7.4 Manual Temperature Setting
If the temperature sensor is not connected, the temperature of the sample solution must be set
manually for accurate Specific Ion or pH measurements. A separate thermometer will be required
for this.
1.
Switch the meter on.
2.
Measure the temperature of the sample.
3.
Select Temperature Calibration (
→
→
F1:Calibrate
F1:Calibrate
→
→
F4:Temperature
F4:Temperature
).
4.
The current temperature setting is now displayed. For example…
Enter Manual Temperature : 25.0
Enter Manual Temperature : 25.0
O
O
C
C
Menu Quits
Menu Quits
5.
Enter the temperature of the sample, using the Numeric Keypad.
Press
to save the new value.
Alternatively, press
to quit and retain the current setting.
6.
The
900-I3
now returns to normal measurement mode. Note the flashing “
M
M
” in the temperature
readout, indicating that Manual Temperature Compensation is in use. For example…
10.0ppM 7.00pH 500.0mV 25.0