8.2 Failed Calibration
If calibration has failed, the GLP function will reset the date and time for the failed parameter to
zero. The 900-I3 still shows the results for the last successful calibration, as shown in the following
examples.
1.
Failed Ion Slope Calibration…
Ion1 Slope=99.0% 00/00/00 00:00
Ion1 Slope=99.0% 00/00/00 00:00
Ion Un-Calibrated F2:Back F4:Next
Ion Un-Calibrated F2:Back F4:Next
2.
Failed pH Asymmetry Calibration…
pH1 Asymmetry=0.10pH 00/00/00 00:00
pH1 Asymmetry=0.10pH 00/00/00 00:00
pH Un-Calibrated F2:Back F4:Next
pH Un-Calibrated F2:Back F4:Next
3.
Failed pH Slope Calibration…
pH1 Slope= 99.0% 00/00/00 00:00
pH1 Slope= 99.0% 00/00/00 00:00
pH Un-Calibrated F2:Back F4:Next
pH Un-Calibrated F2:Back F4:Next
4.
Failed Temperature Offset Calibration…
Temperature Offset=1.0
Temperature Offset=1.0
O
O
c 00/00/00 00:00
c 00/00/00 00:00
Temp. Un-Calibrated F2:Back F4:Next
Temp. Un-Calibrated F2:Back F4:Next
8.3 Printing GLP Information to the RS232 Port
The GLP information stored in the instrument’s memory can be sent to a printer or PC via the
RS232 port. This function is available only when the optional RS232 port is fitted.
1.
Switch the meter on.
2.
Ensure that the
900-I3
RS232 cable is connected to the instrument and to the printer or PC.
3.
Send the GLP information to the RS232 port:
→
→
F4:Setup
F4:Setup
→
→
F3:GLP
F3:GLP
→
→
F3:Print
F3:Print
The message “
Printing GLP Data
Printing GLP Data
” is displayed while sending the data to the RS232 port.
4.
The GLP information is sent to the RS232 port in formatted ASCII text. For example…
900I3 V3.0 R1234 @ 31/12/1998 13:00
900I3 V3.0 R1234 @ 31/12/1998 13:00
Ion1 Offset= 0.1mV @ 31/12/1998 11:00
Ion1 Offset= 0.1mV @ 31/12/1998 11:00
Ion1 Slope= 99.0% @ 31/12/1998 11:05
Ion1 Slope= 99.0% @ 31/12/1998 11:05
pH1 Asy= 0.10pH @ 31/12/1998 11:10
pH1 Asy= 0.10pH @ 31/12/1998 11:10
pH1 Slope= 99.0% @ 31/12/1998 11:15
pH1 Slope= 99.0% @ 31/12/1998 11:15
Ion2 Offset= 0.1mV @ 31/12/1998 11:20
Ion2 Offset= 0.1mV @ 31/12/1998 11:20
Ion2 Slope= 99.0% @ 31/12/1998 11:25
Ion2 Slope= 99.0% @ 31/12/1998 11:25
pH2 Asy= 0.10pH @ 31/12/1998 11:30
pH2 Asy= 0.10pH @ 31/12/1998 11:30
pH2 Slope= 99.0% @ 31/12/1998 11:35
pH2 Slope= 99.0% @ 31/12/1998 11:35
Ion3 Offset= 0.1mV @ 31/12/1998 11:40
Ion3 Offset= 0.1mV @ 31/12/1998 11:40
Ion3 Slope= 99.0% @ 31/12/1998 11:45
Ion3 Slope= 99.0% @ 31/12/1998 11:45
pH3 Asy= 0.10pH @ 31/12/1998 11:50
pH3 Asy= 0.10pH @ 31/12/1998 11:50
pH3 Slope= 99.0% @ 31/12/1998 11:55
pH3 Slope= 99.0% @ 31/12/1998 11:55
Temperature Offset= 1.0oC @ 31/12/1998 12:00
Temperature Offset= 1.0oC @ 31/12/1998 12:00
Ends
Ends