10
3.2 Setting Specific Ion Standards
The Primary and Secondary Specific Ion standards must be set before attempting Specific Ion
calibration.
The Primary and Secondary Specific Ion standard values are entered and stored separately for each
Channel.
The standards for a Channel must be re-entered when changing Specific Ion modes for that
Channel.
All Primary and Secondary Specific Ion standards must be re-entered if the 900-I3 is initialised.
3.2.1
Setting Specific ion standards for ppM, ppK and % readouts
1.
Select the Specific Ion standards set-up menu for Channel 1, 2 or 3.
(e.g.
→
→
F4:Setup
F4:Setup
→
→
F4:Standards
F4:Standards
→
→
F1:Channel 1
F1:Channel 1
).
2.
The Primary Standard set-up screen will now be displayed.
Enter Primary Standard :
Enter Primary Standard : 0
0.00 ppM
.00 ppM
Type in the value of the Primary standard, including the decimal point. Use the
key to
make any corrections.
Press
to save the Primary standard.
Alternatively, press
to quit without changing the current setting.
3.
The Secondary Standard set-up screen is now displayed.
Enter Secondary Standard :
Enter Secondary Standard : 0
0.00 ppM
.00 ppM
Type in the value of the Secondary standard, including the decimal point. Use the
key to
make any corrections.
The Secondary Standard must be at least 2 to 100 times higher or lower than the primary
standard.
Press
to save the Secondary standard.
Alternatively, press
to quit without changing the current setting.