! If the problem persists, contact Frontier Therapeutics Ltd Tel. +44 (0) 1495 235800 for
assistance. Do not open the Control Unit. Opening the unit could cause personal injury or
equipment damage.
CPR Emergency Deflation System
CPR Deflation Valve Open
The Toto lateral turning system features a quick release CPR deflation valve which allows for rapid
deflation for emergency procedures.
The CPR deflation valve can be found on the air hoses between the Toto Touch Control Unit and
the platform, clearly marked with a yellow tab.
In the event of an emergency, firmly pull and release the yellow tab.
To re-inflate the system after the CPR tag has been removed, replace the CPR tag ensuring both
sealing connectors are firmly attached, and restart the Toto Touch Control Unit.
All five
Inflate Time
LEDs are flashing, an
audible beep can be
heard.
(E - I)
•
Air Flow between the
Control Unit and platform
is obstructed.
1. Press the Alarm Mute button.
2. Check for twists or kinks in the air hoses between
the platform and Control Unit.
3. Resolve and wait for the alarm to reset.
! If the problem persists, contact Frontier
Therapeutics Ltd Tel. +44 (0) 1495 235800 for
assistance.
Service LED
permanently lit
Amber.
(O)
•
Service required.
•
Indicates the need of periodic maintenance.
Please contact Frontier Therapeutics Customer
Services for support Tel. +44 (0) 1495 235800
! The Control Unit will continue to function even
though the service LED remains lit.
Interface panel is lit
but unresponsive.
•
Interface panel is locked.
•
Press and hold ALARM MUTE button for 4 seconds.
Audible beep, no
LEDs lit.
•
Accidental removal of
power cord.
•
Accidental mains wall
power turned off.
•
Mains power failure.
•
Reconnect mains power lead cable to Control Unit.
•
Check mains power wall switch is turned on.
CPR Deflation Valve Closed
CPR Deflation Valve Open
System Error Trouble Shooting Guide
System Error
Indicator
Cause
Initial Checks / Trouble Shooting
No LEDs lit and no
audible beeps can be
heard.
•
Control Unit may not
be attached to a power
source.
•
Fuse may need replacing.
1. Check the Control Unit is connected to a mains
power outlet and is switched on.
2. Check the Control Unit is switched on.
3. Switch off, unplug the Control Unit:
- Check the mains plug fuse (3 AMP).
- Check Control Unit fuse (1 AMP).
4. Restart Control Unit.
! If the problem persists, contact Frontier
Therapeutics Ltd Tel. +44 (0) 1495 235800 for
assistance.
!
Do open the Control Unit. Opening the unit could
cause personal injury or equipment damage.
All five
Rest Time
LEDs are flashing, an
audible beep can be
heard.
(J - N)
•
Pressure has fallen below
the minimum operational
requirements.
1. Press the Alarm Mute button.
2. Check the CPR valve is properly closed, ensuring all
sealing connectors are fully pressed in.
3. Check each air cell is securely attached to the
connecting air cell hose.
4. Check all air cells, and hoses for any air leakage.
! If the problem persists, contact Frontier
Therapeutics Ltd Tel. +44 (0) 1495 235800 for
assistance.