background image

6

7

Intended Use

Product Description

Toto is an automated lateral turning system which consists of a Toto Touch Control Unit and a 
turning platform with cover.

Designed to assist with patient turning and pressure ulcer prevention plans, the system is based 
on inflating alternating air cells within the platform. These turn patients laterally via the mattress 
surface to a maximum of either 30 or 45 degrees depending on which version is used.

The digital Control Unit provides customised care with a user determined side selection feature 
and accurately timed digital inflation and rest durations. Sophisticated blockage and air leak alarms 
help ensure patient safety, as well as an interface lock to avoid unwanted setting changes. 

A convenient therapy pause mode enables patient interaction without having to stop therapy and 
a service indicator to maintain performance also comes as standard.

The system is compatible with profiling beds and both standard and alternating pressure relieving 
equipment. It is also compatible with domestic mattresses, except pocket sprung.

Toto lateral turning system does not replace pressure relieving and redistribution surfaces and is 
supplementary to their use.

Indications

Toto lateral turning system is suitable for the prevention of pressure ulcers for the following 
patients:

•  Those identified as being at moderate to high risk of developing pressure ulcers as defined by 

commonly used assessment tools.

•  Weighing up to 190kg (419lbs).

•  Who are unable to change their position without assistance.

•  Identified as requiring regular turning and who are non-compliant with manual turning schedules.

Contraindications

•  Do not use without a suitable pressure redistribution surface & package of care.

•  Do not use with patients with unstable spinal cord injuries or who are undergoing traction.

•  Do not use with patients with equilibrium disturbances.

•  Not to be used where patients must be supported by wedges or pillows etc.

Intended Care Setting

Toto Lateral turning system is intended to be used in the following care settings: 

•  Long term or extended care, such as nursing homes and care homes.

•  Home care.

•  Hospitals.

Safety

 !  Warnings and Precautions

Individual patient assessments should be made prior to use and appropriate pressure relieving 
devices put in place. Please note the following:

•  Caution is advised when using with patients with underlying neurological disorders which may 

lead to increased anxiety.

•  Unusual body shapes and body postures.

•  Other medical devices should be assessed for suitability.

•  The Toto lateral turning system should be fully deflated when attempting to leave the bed.

•  Ensure that the Toto lateral turning system is discontinued or therapy pause mode is engaged 

for feeding times to avoid the risk of spillages.

•  During hygiene procedures, please use therapy pause mode.

•  When elevating a profiling bed back rest, raise the knee break first.

•  Ensure that the surface of the bed frame is clear from debris before placing the Toto lateral 

turning system platform, as this can damage the coating of the cover.

•  The system should not be used without a mattress.

•  Do not strap the mattress to the bed frame as this will prevent Toto lateral turning system from 

operating.

•  If mattress overlays are being used, these must be securely fastened to the mattress, not the 

Toto lateral turning system platform.

•  The Toto lateral turning system can be used with positioning systems, bed levers and linen 

cradles, if their support frame fits horizontally across the bed. These must be positioned under 
the Toto lateral turning system platform.

•  Evacuation hammocks must be placed between the Toto lateral turning system platform and 

the mattress.

•  If using bed sides, check for clearance from the top of the mattress to the bottom of the bed 

side (minimum of 185mm) and assess for risk of entrapment.

•  If using foot board protectors, assess for compatibility.

•  Do not place the Control Unit directly underneath the bed frame if placing on the floor to avoid 

lowering the bed onto the Control Unit.

•  Do not use if mains power lead or plug is damaged.

Содержание GP0133-A50-01

Страница 1: ...Toto Lateral Turning System with Toto Touch Control Unit User Guide www frontier group co uk...

Страница 2: ...wed Do not pierce or cut Do not iron Line dry Refer to user manual CE Compliant Type BF applied part Keep dry Class II equipment Disposal do not dispose of this product as unsorted domestic waste It m...

Страница 3: ...lease Pull Tag 45 XL 35 1 4 271 11 Quick Release Pull Tag 30 Product Weight kg lbs Maximum User Weight kg lbs Width mm Inches Length mm Inches Standard 18 40 190 419 745 29 1850 73 45 18 40 190 419 74...

Страница 4: ...es or pillows etc Intended Care Setting Toto Lateral turning system is intended to be used in the following care settings Long term or extended care such as nursing homes and care homes Home care Hosp...

Страница 5: ...the Control Unit Black to Control Unit ensuring the hoses are not twisted or kinked F 10 Connect the Control Unit to the mains power via the cable supplied G A The platform should be placed directly o...

Страница 6: ...d hold the start stop button for 3 seconds A single beep will confirm activation and the start stop green LED will be lit Note the system default setting will automatically be selected o Air cell infl...

Страница 7: ...utton a single beep will confirm activation The start stop LED will flash when active and all other LED s will turn off After 10 minutes in Therapy Pause Mode the alarm will sound To continue in this...

Страница 8: ...dible beep and Low system pressure Pressure in the system has fallen below the minimum operational requirements Five Rest time LEDs flashing when activated High system pressure Air flow between the Co...

Страница 9: ...ion even though the service LED remains lit Interface panel is lit but unresponsive Interface panel is locked Press and hold ALARM MUTE button for 4 seconds Audible beep no LEDs lit Accidental removal...

Страница 10: ...on and feed air cell hoses through apertures at the bottom of the cover 8 Reattach the hoses using the quick connectors Red to Red ensuring they are not twisted or kinked Disposal IMPORTANT In order t...

Страница 11: ...n may be achieved by laundering at temperatures not exceeding 65 C for 10 minutes or 73 C for three minutes 1 1 Protect Rinse and Dry BHTA guidance on the care cleaning and inspection of healthcare ma...

Страница 12: ...ot affect your statutory rights Fire Testing The Toto lateral turning system platform meets the requirements of The Furniture Furnishings Fire Safety Regulations 1988 Meeting the requirements of BS 71...

Страница 13: ...rk of Frontier Therapeutics Ltd November 2017 Frontier Therapeutics Limited All rights reserved GP0133_A50_01 w ww your turn org u k Frontier Therapeutics Limited Newbridge Road Industrial Estate Blac...