Instructions for Use
Platform Installation and Operating Guide
To ensure the system is correctly installed and functions smoothly, please follow this step-by-step
guide.
1. Remove all items from the packaging and inspect for damage.
2. The platform should be placed directly onto the bed frame and underneath the existing mattress.
DO NOT strap or fasten the mattress to the bed frame. (A)
3. Inspect the platform and cover for any visible signs of damage. (B)
4. Ensure the air cells within the platform are secured in position via the Velcro®.
5. Fit the cover and close zip, ensuring the air cell hoses are fed through the apertures at the foot
end of the cover.
Note: The platform cover should be free to move, allowing each side of the platform unrestricted
room to inflate.
6. Using the quick connectors, connect the air inflation hoses to the air cell hoses - 'Red to Red' -
ensuring the hoses are not twisted or kinked. (C)
7. Check the CPR deflation valves are securely in place to ensure no air leaks. (D)
8. Using the bed hooks, hang the Control Unit on footboard of the bed. (E)
9. Using the quick connectors, connect the air inflation hoses to the Control Unit - 'Black to Control
Unit' - ensuring the hoses are not twisted or kinked. (F)
10. Connect the Control Unit to the mains power via the cable supplied. (G)
A.
The platform should be placed directly
onto the bed frame and underneath the
existing mattress.
B.
Inspect the platform for visible signs of
damage.
C.
Using the quick connectors, connect
the air inflation hoses to the air cell hoses
- 'Red to Red' - ensuring the hoses are
not twisted or kinked.
D.
Check the CPR deflation valves are
securely in place to ensure no air leaks.
E
. Using the bed hooks, hang the Control
Unit on footboard of the bed.
F.
Using the quick connectors, connect
the air inflation hoses to the Control Unit
'Black to Control Unit' - ensuring the
hoses are not twisted or kinked.
G.
Connect the Control Unit to the mains
power via the cable supplied.
Patient Positioning
Patients should be positioned along the centre line of the mattress by aligning the patients nose
with the centre point of the head board. Ensure head and shoulders are supported to prevent neck
roll and pivoting at the hips during use.
For patients lying on their side, if they can self-support and they are in the middle of the bed, Toto
lateral turning system can be used.
If patients cannot control their limbs and bed sides are being used, padding or net infills are
recommended.
Centre Line