Instructions for Use

Platform Installation and Operating Guide

To ensure the system is correctly installed and functions smoothly, please follow this step-by-step 
guide.

1.  Remove all items from the packaging and inspect for damage.

2.  The platform should be placed directly onto the bed frame and underneath the existing mattress.

DO NOT strap or fasten the mattress to the bed frame. (A)

3.  Inspect the platform and cover for any visible signs of damage. (B)

4.  Ensure the air cells within the platform are secured in position via the Velcro®.

5.  Fit the cover and close zip, ensuring the air cell hoses are fed through the apertures at the foot 

end of the cover.

Note: The platform cover should be free to move, allowing each side of the platform unrestricted 
room to inflate.

6.  Using the quick connectors, connect the air inflation hoses to the air cell hoses - 'Red to Red' - 

ensuring the hoses are not twisted or kinked. (C)

7.  Check the CPR deflation valves are securely in place to ensure no air leaks. (D)

8.  Using the bed hooks, hang the Control Unit on footboard of the bed. (E)

9.  Using the quick connectors, connect the air inflation hoses to the Control Unit - 'Black to Control 

Unit' - ensuring the hoses are not twisted or kinked. (F)

10. Connect the Control Unit to the mains power via the cable supplied. (G)

A. 

The platform should be placed directly 

onto the bed frame and underneath the 
existing mattress.

B. 

Inspect the platform for visible signs of 

damage. 

C.

 Using the quick connectors, connect 

the air inflation hoses to the air cell hoses 
- 'Red to Red' - ensuring the hoses are 
not twisted or kinked.

D.

 Check the CPR deflation valves are 

securely in place to ensure no air leaks.

E

. Using the bed hooks, hang the Control 

Unit on footboard of the bed.

F.

 Using the quick connectors, connect 

the air inflation hoses to the Control Unit 
'Black to Control Unit' - ensuring the 
hoses are not twisted or kinked.

G. 

Connect the Control Unit to the mains 

power via the cable supplied.

Patient Positioning

Patients should be positioned along the centre line of the mattress by aligning the patients nose 
with the centre point of the head board. Ensure head and shoulders are supported to prevent neck 
roll and pivoting at the hips during use.

For patients lying on their side, if they can self-support and they are in the middle of the bed, Toto 
lateral turning system can be used.

If patients cannot control their limbs and bed sides are being used, padding or net infills are 
recommended.

Centre Line

Содержание GP0133-A50-01

Страница 1: ...Toto Lateral Turning System with Toto Touch Control Unit User Guide www frontier group co uk...

Страница 2: ...wed Do not pierce or cut Do not iron Line dry Refer to user manual CE Compliant Type BF applied part Keep dry Class II equipment Disposal do not dispose of this product as unsorted domestic waste It m...

Страница 3: ...lease Pull Tag 45 XL 35 1 4 271 11 Quick Release Pull Tag 30 Product Weight kg lbs Maximum User Weight kg lbs Width mm Inches Length mm Inches Standard 18 40 190 419 745 29 1850 73 45 18 40 190 419 74...

Страница 4: ...es or pillows etc Intended Care Setting Toto Lateral turning system is intended to be used in the following care settings Long term or extended care such as nursing homes and care homes Home care Hosp...

Страница 5: ...the Control Unit Black to Control Unit ensuring the hoses are not twisted or kinked F 10 Connect the Control Unit to the mains power via the cable supplied G A The platform should be placed directly o...

Страница 6: ...d hold the start stop button for 3 seconds A single beep will confirm activation and the start stop green LED will be lit Note the system default setting will automatically be selected o Air cell infl...

Страница 7: ...utton a single beep will confirm activation The start stop LED will flash when active and all other LED s will turn off After 10 minutes in Therapy Pause Mode the alarm will sound To continue in this...

Страница 8: ...dible beep and Low system pressure Pressure in the system has fallen below the minimum operational requirements Five Rest time LEDs flashing when activated High system pressure Air flow between the Co...

Страница 9: ...ion even though the service LED remains lit Interface panel is lit but unresponsive Interface panel is locked Press and hold ALARM MUTE button for 4 seconds Audible beep no LEDs lit Accidental removal...

Страница 10: ...on and feed air cell hoses through apertures at the bottom of the cover 8 Reattach the hoses using the quick connectors Red to Red ensuring they are not twisted or kinked Disposal IMPORTANT In order t...

Страница 11: ...n may be achieved by laundering at temperatures not exceeding 65 C for 10 minutes or 73 C for three minutes 1 1 Protect Rinse and Dry BHTA guidance on the care cleaning and inspection of healthcare ma...

Страница 12: ...ot affect your statutory rights Fire Testing The Toto lateral turning system platform meets the requirements of The Furniture Furnishings Fire Safety Regulations 1988 Meeting the requirements of BS 71...

Страница 13: ...rk of Frontier Therapeutics Ltd November 2017 Frontier Therapeutics Limited All rights reserved GP0133_A50_01 w ww your turn org u k Frontier Therapeutics Limited Newbridge Road Industrial Estate Blac...

Отзывы: