When installing the front lens unit, make sure that the front lens unit fixing lever
is fixed securely at the LOC side.
An injury may be caused by the front lens unit falling off.
57
Before setting the front lens, move the microscope at least 200mm upwards.
Otherwise, the components may come into contact with each other and cause
an injury.
70
Adjust the front lens and the treatment section, taking care not to hit the patient
with the front lens.
This could injure the patient.
70
When setting/storing the front lens unit, make sure that the front lens unit fixing
lever is fixed securely at the LOC side.
An injury may be caused by the front lens unit falling off.
70
Make sure that the front lens unit is securely attached to the optical unit after
connecting the two.
The front lens may move suddenly, causing an injury.
70
Before moving the instrument, make sure that no-one is near. Then, move it
carefully.
Injury could be caused by the instrument touching anything.
83
85
Before using this instrument, make sure that the handles, levers, knobs and
rings with red marks are securely tightened.
Any of these falling off could cause injury or even death.
103
Make sure no error code is displayed in the light intensity display window.
If errors are displayed, the instrument may not operate normally, causing
problems during an operation.
103
While the anterior eye section observation lens is in use, do not operate the
front lens IN/OUT lever and the front lens connecting/disconnecting knob. The
anterior eye section observation lens may fall off, causing an injury.
77
While the peripheral observation prism is in use, do not operate the front lens
IN/OUT lever and the front lens connecting/disconnecting knob.
The peripheral observation prism may fall off, causing an injury.
77