36
PREPARATIONS BEFORE USE
7
Adjust the balance of the 2nd arm.
After installing or removing the accessories, the 2nd arm balance must be readjusted.
Refer to "ADJUSTING THE 2ND ARM BALANCE: OFFISS, OFFISS Lite, Pro, Pro Lite" on
P.41.
SETTING THE ACCESSORIES: Standard
1
While loosening the 2nd arm vertical movement fixing handle, hold the microscope operation
handle and move the 2nd arm to the upper limit position.
WARNING
Do not install/remove the accessories above the patient. An accessory
falling off could cause injury or even death.
WARNING
Before installing the accessories, make sure all the arms are securely
locked. The 2nd arm may move suddenly, causing an injury.
WARNING
After installing/removing the accessories, make sure that handles,
levers, knobs and rings are securely tightened. Any of these falling off
could cause injury or even death.
WARNING
Before using this instrument, adjust the balance of the 2nd arm. The 2nd
arm may move up and down suddenly during an operation, causing an
injury.
WARNING
Release the 2nd arm lower limit lock while holding it at the end. The 2nd
arm may move up and down suddenly, causing an injury.
CAUTION
Do not open the instrument, as this may lead to an electric shock.
NOTE
Ask your distributor or the Topcon offices stated on the back cover to repair the
instrument.