PREPARATIONS BEFORE USE
PREPARATIONS BEFORE USE
Before using this instrument, carry out the daily check. (Refer to "DAILY CHECK" on P.103.)
STERILIZED CAP: Common
Sterilize caps as follows:
Method: High pressure steam sterilization
Condition: Leave cap in the saturated steam (134±4°C) for 20 minutes.
PREPARATIONS FOR STERILIZING THE FRONT LENS UNIT: OFFISS, OFFISS Lite
Place the front lens unit, peripheral prism and anterior eye section observation lens in the sterilized tray
and sterilize them as follows:
Method: EOG (ethylene oxide gas) sterilization
Condition: Make sure the devices used are properly sterilized.
TURNING ON THE POWER: Common
1
Connect the power plug to a wall-outlet.
2
Switch the power switch "
" (ON).
On the light intensity display window, the version information of the software installed in this
instrument is displayed for several seconds. (Display example: 1.03) If the software of
Version 1.02 or previous is installed, the version information will not be displayed.
3
Make sure that the power indicator lights up green.
4
Make sure that no error code is displayed on the light intensity display window. (Refer to
"ERROR CODES" on P.88.)
WARNING
Make sure no-one is too close to the instrument before moving the arm.
Anyone touching the instrument may be injured.
NOTE
Make sure the electricity is properly grounded.