99
MAINTENANCE AND INSPECTION: Common
The capacity of the circuit breaker on the primary side is 5A. "5A" is indicated at the center.
1
When the circuit breaker is operating, make sure that the breaker end pops out.
2
Turn off the power switch.
3
Push in the circuit breaker end.
4
Turn on the power switch again.
NOTE
If the circuit breaker does not operate after resetting, the instrument may have a
problem. Contact your distributor or the Topcon offices stated on the back cover.
NOTE
Even if the instrument operates normally after resetting the circuit breaker,
contact your distributor or the Topcon offices stated on the back cover.