SPECIFICATIONS AND PERFORMANCE
ELECTROMAGNETIC COMPATIBILITY
The product conforms to the EMC standard (IEC 60601-1-2 Ed3.0:2007)
a)MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information provided in the
ACCOMPANYING DOCUMENTS.
b)Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
c)The use of ACCESSORIES, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the EQUIPMENT or SYS-
TEM as replacement parts for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the EQUIPMENT or SYSTEM.
d)The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equip-
ment. If adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be
observed to verify normal operation in the configuration in which it will be used.
e)The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other
than those specified may result in increased EMISSION or decreased IMMUNITY of the
EQUIPMENT or SYSTEM.
Guidance and manufacturer's declaration - electromagnetic emissions
The KR-1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the KR-1 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The KR-1 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The KR-1 is suitable for use in all establishments
other than domestic and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic pur-
poses.
Harmonic emissions
IEC61000-3-2
Complies
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies