15
COMPONENTS
PRINTER OUTPUT
KRT typical value style and KRT print format are HV
The reliability factor is defined with integers 1 to 9 in increasing order of reliability.
Additionally, if the reliability is high enough, the reliability factor is not shown on the
printout.
The Near PD value is calculated based on the ADD.
( ) appears when normal measurement is not expected due to eyelid, eyelash, or
blinking.
*-mark appears when normal measurement is not expected with the
button selected.
Operator ID
Device ID number
3 readings of REF right
measurement (recordable
up to 10 measurements)
Pupil distance (PD value)
VD (vertex distance)
Work ID No.
Bar code
Patient No.(Patient ID
when patient ID is input)
Serial number
Cylinder sign
Typical value of right eye
SPHERICAL EQUIVALENT
of right eye
The ( ) mark is added when
measurement values are not
fully reliable.
ADD
(standard additional power)
Near vision PD value
Refractive power measurement result (L)
Refractive power measurement result (R)
Typical measured value of right eye corneal curvature
Typical measured value of
Left eye corneal curvature
Measured value of horizontal
corneal curvature
Measured value of vertical corneal
curvature
3 readings of kerato-cylinder value,
average value and kerato-cylinder
value (recordable up to 10 mea-
surements each for right/left eye)
Left eye
corneal curvature measurement
Measured value of right eye cornea
Corneal astigmatic axis angle
Peripheral KRT measurement
values near 6mm
Peripheral KRT value
Typical measured value of
Peripheral KRT value (R)
C(Cataract mode) mark
Reliability factor
Cataract
Содержание KR-1
Страница 2: ......
Страница 76: ...KR 1 AUTO KERATO REFRACTOMETER 41844 95015 Printed in Japan 2012 09 100LW5 BACK COVER ...