48
SETTING FUNCTIONS ON SETUP SCREEN
LIST OF SETUP ITEMS
Setup items are categorized into 6 large indexes.
"Initial" .....................items related to the initial status after power on
"Print" ......................items related to output from the internal printer
"Comm" ...................items related to data output with the external device
"LAN".......................items related to output using the LAN
"Operator ID" ...........items related to Operator ID.
"Special" ..................items related to maintenance (for service engineer only)
INITIAL (INITIAL SETTING)
Initial contains settings related to the initial status after power on, clearing all measure-
ment values, etc.
Descriptions
Options
Details
Initial value
Buzzer
OFF
Buzzer does not sound.
ON
ON
Buzzer sounds.
Start mode
MANUAL
Default measurement mode is MANUAL.
AUTO
AUTO
Default measurement mode is AUTO.
Auto print
OFF
Not printed automatically.
ON
ON
After AUTO measurement,
results are printed out automatically.
Printer
OFF
Internal printer is disabled.
ON
ON
Internal printer is active.
Patient No. reset
OFF
Patient No. is not reset upon power on.
OFF
ON
Patient No. is reset upon power on.
Show patient ID
OFF
Patient ID is not displayed.
ON
ON
Patient ID is displayed.
Required
patient ID
OFF
Patient ID is not displayed.
OFF
ON
Patient ID is displayed.
Device ID
number
1-99
Set by ten-key display.
Sets the Device ID number.
1
Show Device ID
number
OFF
Device ID is not required.
OFF
ON
Device ID is required.
Stand by mode
RIGHT
Waiting at the initial position for right eye measurement.
RIGHT
LEFT
Waiting at the initial position for left eye measurement.
LAST
Waiting at the last position of the measured eye.
Auto power save
OFF
Power save function is not used.
10min
1min
Power save status in 1min after last operation.
5min
Power save status in 5min after last operation.
10min
Power save status in 10min after last operation.
20min
Power save status in 20min after last operation.
30min
Power save status in 30min after last operation.
60min
Power save status in 60min after last operation.
Cont. Cycle
1-10
Set by ten-key display
The number of continuous measurements.
3
Add Measure
0-99
Set by ten-key display.
When the measurement is error,
set the number of times of remeasurement.
1
Continuous fog
Every time
Continuous fog is applied every time.
Once
Once
Continuous fog is applied only once
before the 1st measurement.
Date/Time
Set by ten-key display.
Sets year, month, day, time (24hrs), minute and second.
Installation
date/time
Sph/Cyl step
0.12
Sph/Cyl is displayed by 0.12D step.
0.25
0.25
Sph/Cyl is displayed by 0.25D step.