OBJECTIVE OPERATIONS
Photography screen (when automatic photography has failed)
If an error message is displayed during automatic photography, perform adjustment by manual photog-
raphy. Then, press the capture start button to continue the photography.
• If an error has occurred during automatic alignment to pupil:
If the patient's face or eye cannot be detected or the pupil center is not tracked, the following mes-
sages are displayed.
Perform one of the following operations.
- Set the patient properly, press the [Retry] button and perform automatic alignment again
according to the message.
- Press the [Manual] button to shift to the manual adjustment mode.
After tapping the pupil in the movie area, press the capture start button to continue the photogra-
phy.
• If an error has occurred during automatic focus and automatic optimizing:
If automatic photography has failed, shift to the manual adjustment mode. After adjusting, press
the capture start button to continue the photography.
When the eyelid is not opened fully
When the eye fixation is not stabilized