116
REFERENCE MATERIAL
3D Scan
Move on the inside of the square, which is composed of the given start point and end point, horizontally
and vertically by the step divided by the given resolution.
The scan for [3D Macula] and [3D Disc] is shown below. ("6.0×6.0mm" is initially set for the scan length
in both of these scans.)
The scan for [3D Macula (V)] and [3D Wide] is shown below. (As the scan length, "7.0×7.0mm" is fixed
for [3D Macula (V)] and "12.0×9.0mm" is fixed for [3D Wide].)
Radial Scan
In the scan range, perform scanning by the specified diameter and by the step divided by the given res-
olution. "6.0mm" is fixed for the scan length. The start point for Line-Scan and rotating direction are
reversed for each of right and left eyes. For the right eye, rotation is done counterclockwise in the hor-
izontal direction. For the left eye, rotation is done clockwise in the horizontal direction.
[3D-Scan]
End point
Live-Scan
position
Start point
End point
Start point
Live-Scan
position
Left eye
Right eye
S
S
I
I
N
N
T
T
Start point
End point
Live-Scan
position
Start point
End point
[3D-Scan](V)
Left eye
Right eye
S
S
I
I
N
N
T
T
Live-Scan
position
1
6
5
4
3
2
1
2
3
4
5
6
Left eye
Right eye
S
I
N
T
S
I
N
T
Содержание 3D OCT-1
Страница 1: ...USER MANUAL for instrument body 3D OPTICAL COHERENCE TOMOGRAPHY 3D OCT 1 Version 1 2x...
Страница 2: ......
Страница 67: ...65 OBJECTIVE OPERATIONS 2 After the photograph is taken the result is automatically displayed 1234ABCD...
Страница 82: ...80 DETAILS OF THE SETTING MENU 3 Make sure that the icon is added...
Страница 126: ...3D OPTICAL COHERENCE TOMOGRAPHY 3D OCT 1 47010 91942 Printed in Japan 1307 100TH 2...