SPECIFICATIONS AND PERFORMANCE
Guidance and manufacturer's declaration - electromagnetic immunity
The 3D OCT-1 is intended for use in the electromagnetic environment specified below. The customer
or the user of the 3D OCT-1 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Main power quality should be that of a
typical commercial or hospital environ-
ment.
Surge
IEC 61000-4-5
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth
± 1 kV
line(s) to line(s)
± 2 kV
line(s) to earth
Main power quality should be that of a
typical commercial or hospital environ-
ment.
Voltage dips, short
interruptions and
Voltage variations
on power supply
input lines
IEC 61000-4-11
<5%
U
t
(>95% dip in
U
t
)
for 0.5 cycle
40%
U
t
(60% dip in
U
t
)
for 5 cycles
70%
U
t
(30% dip in
U
t
)
for 25 cycles
<5%
U
t
(>95% dip in
U
t
)
for 5 sec
<5%
U
t
(>95% dip in
U
t
)
for 0.5 cycle
40%
U
t
(60% dip in
U
t
)
for 5 cycles
70%
U
t
(30% dip in
U
t
)
for 25 cycles
<5%
U
t
(>95% dip in
U
t
)
for 5 sec
Main power quality should be that of a
typical commercial or hospital environ-
ment. If the user or the 3D OCT-1
requires continued operation during
main power interruptions, it is recom-
mended that the 3D OCT-1 be powered
from an uninterruptible power supply or
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE
U
t
is the a.c. main voltage prior to application of the test level.
Содержание 3D OCT-1
Страница 1: ...USER MANUAL for instrument body 3D OPTICAL COHERENCE TOMOGRAPHY 3D OCT 1 Version 1 2x...
Страница 2: ......
Страница 67: ...65 OBJECTIVE OPERATIONS 2 After the photograph is taken the result is automatically displayed 1234ABCD...
Страница 82: ...80 DETAILS OF THE SETTING MENU 3 Make sure that the icon is added...
Страница 126: ...3D OPTICAL COHERENCE TOMOGRAPHY 3D OCT 1 47010 91942 Printed in Japan 1307 100TH 2...