TH2523 Series Operation Manual
Ver1.2
23
Figure 3-8 Statistic Display Page
The function of this page is to count measurement data, such as some engineering coefficients,
the PASS/FAIL rate. In this page, it is also workable to analyze the average value of results for
multiple measurements. Details are as below.
3.7.1
Mode
Move the cursor to it and select different modes through the soft key. There are two modes for
comparison of high-low value: ABS (resistance value, inductance value and voltage value of
Hi/Lo) and % (percentage of nominal value and Hi/Lo).
NOTE: the formula of percentage mode:
Upper limit value= nominal value*(1+ upper limit%)
Lower limit value= nominal value*(1+ lower limit%)
3.7.2
NOMA/NOMB
Move the cursor to it and input the data through the numerical keyboard.
A: nominal value of primary parameter
B: nominal value of secondary parameter
3.7.3
Number
Move the cursor to it and input the total measurement times of statistic analyses through the