— 33 —
5-6. Setting Operating Parameters
CAUTION
Do not exceed the maximum allowable RCF of the tube.
●
Failure to do so may break the tube and cause the rotor to fall off during
spinning.
When the mass density of the sample exceeds 1.2 g/cm
3
, do not
centrifuge at speed exceeding the maximum allowable speed of the
rotor which is respectively calculated.
●
An excessive speed may damage the rotor, bucket or rack, thereby causing
the rotor to fall off during spinning.
Press the key. A single "pip" sounds.
The temperature display switches from the actual value to the set value and begins to flash.
Turn the Jog Dial to set a desired temperature.
<
To set the temperature
>
①
②
※
Temperature setting range : -9℃ to 35℃ (in one degree increments)
※
Set temperatures referring to "11-2. Cooling Capacity" on page in this Operator's Manual for
controllable temperature range.
After setting the temperature, leave it for approximately two seconds to finalize the setting. The
temperature display stops flashing and returns to the actual value from the set value.
③
※
When turning the power on, the temperature is set to the last operating value.
TEMP
Press the key. A single "pip" sounds.
The speed/RCF display begins to flash.
Check that the "rpm" lamp of the rpm/G unit indicators lights up and turn the Jog Dial to set a
desired speed.
<
To set the speed
>
①
②
※
The rpm/G unit lamps light up alternately by pressing the key.
※
Speed setting range MX-307/207 : 300
〜
16,000rpm
MX-107 : 300
〜
15,000rpm
After setting the speed, leave it for approximately two seconds to finalize the setting.
③
SPEED/RCF
SPEED/RCF