![Tomy MX-107 Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/tomy/mx-107/mx-107_operators-manual_1126195042.webp)
— 41 —
6-5. Memory Function
The MX-307, 207 and 107 have a memory function which allows storing frequently used settings
of operating parameters. Two sets of operating parameters including temperature, speed/RCF,
run time and Soft Brake setting can be stored in memory.
Make sure that the rotor stops.
Set each value for the operating parameter to be stored, referring to "5-6. Setting Operating
Parameters" in this Operator's Manual.
Press and hold down the desired memory number key ( or )
for longer than three seconds. A beep sounds and the indicator lamp on the selected memory
key lights up to show the memory setting has been achieved.
※
Please note that the previous operating parameters in the selected memory have been cleared.
①
②
③
<
To store operating parameters
>
MEMORY2
MEMORY1
Make sure that the rotor stops.
Press the desired memory number key ( or ).
The indicator lamp on the selected memory key lights up to show the desired set of operating
parameters is recalled and appears on the display.
※
Please note that the recalled operating parameters can be modified for a particular operation, if
necessary, however, the stored data in the memory will not be changed.
※
To change the stored set of operating parameters, take the necessary steps according to the instruction,
<To store operating parameters> stated above.
①
②
<
To recall the stored operating conditions
>
MEMORY2
MEMORY1
Содержание MX-107
Страница 19: ... 18 MEMO ...
Страница 49: ... 48 Disinfect the unit with ethanol To disinfect the main unit ...
Страница 57: ... 56 MEMO ...
Страница 63: ... 62 MEMO ...
Страница 68: ... 67 MEMO ...
Страница 83: ... 82 MEMO ...