SLIT LAMP TSL-7000Z by TOMEY

SLIT LAMP TSL-7000Z by TOMEY

6.  INSTRUCTIONS FOR USE

6.1  SETTING THE BINOCULARS

It is vital that the binoculars are optimised for the user’s optical 

correction in order to obtain focused binocular images.

1.  Remove the Test Bar locating hole cover plate (16) and place the test bar focus in the test 

bar location hole at the base of the microscope arm. To access the location hole fi rst remove 

the cover. The test bar should be set with the fl at projection face towards the Slit Lamp 

Microscope. The illumination and the microscope should be in the zero degrees position.

2.  Turn on the Slit Lamp and set the slit to full width (14), set the magnifi cation to x16 (27).
3.  Adjust the eyepieces pupillary distance by holding both eyepiece bodies and rotating 

them inwards or outwards until they are correct for your PD.

4.  Turn both eyepieces (28) to maximum plus (+) 

correction.

5.  Close one eye, and with the other eye look through 

the microscope slowly turning the open eye 

eyepiece towards the minus (-) position until the 

image of the test bar is in focus. Stop.

6.  Repeat the above process for the other eyepiece.
7.  Make a note of the positions of the eyepieces so that you can set them quickly if the Slit 

Lamp has been used by another clinician.

8.  Note – younger examiners are recommended to compensate for their ability to 

accommodate by further adjusting the eyepieces by minus one (-1) or minus two (-2) 

dioptres.

16

 

27

28

6.2  PREPARING THE PATIENT AND USING THE SLIT LAMP

Parts of the slit lamp coming in to contact with the patient should be 

cleaned in accordance with these instructions prior to the examination. 

Tomey recommends the use of disposable hygienic chinrest tissues on the 

Chinrest before the patients place their chin on it.
Never use the instrument if visibly damaged and periodically inspect it for 

signs of damage or misuse.

1.  The patient should be as comfortable as possible and with the patient in the chinrest 

adjust the chinrest height (6) so that the patient’s eyes are level with the height marking 

(3) on the chinrest support.

2.  Focus the eyepieces using the test bar as described earlier, and if you have not already 

done so set them to your interpupillary distance by holding both eyepiece bodies and 

rotating them inwards or outwards until they are correct for your PD.

3.  Switch on the illumination, making sure the rheostat (21) is set to a low level to 

minimise the patient’s exposure to light hazard.

4.  Rotate the joystick (20) until the light beam is at eye level.
5.  Holding the joystick vertical, move the slit Lamp base towards the patient until the slit 

beam appears focused on the patient’s cornea.

6.  Adjust slit width (14), magnifi cation (27), slit rotation (13) and slit angle etc. as required 

to perform the examination.

7.  To offset the slit for sclerotic scatter or retro illumination, rotate the illumination prism 

assembly (11) to the left or right as desired.

8.  When using the blue fi lter (12) the user may wish to insert the yellow barrier fi lter (9). 

The yellow barrier fi lter is out when the knob is up, in when it is down.

9.  When the examination is complete, set the rheostat to a low level and switch off the Slit 

Lamp.

Shut down after every use. In case the dust cover is used: risk of 

overheating.

20

21

9

12

27

11

13

14

3

31

Содержание TSL-7000Z

Страница 1: ...TSL 7000Z SLIT LAMP OPERATOR MANUAL Z Series...

Страница 2: ...IONS FOR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYSTEMS 1...

Страница 3: ...f the anterior eye segment The Tomey Slit Lamp is an instrument consisting of a light source that can be focused to shine a thin sheet slit of light into the eye It is used in conjunction with a biomi...

Страница 4: ...is disconnected LEDs can reach high temperatures in use allow to cool before handling Do not exceed maximum recommended exposure time Should the instrument suffer shocks for example should it accident...

Страница 5: ...f used cleaning materials 4 NAMES OF CONTROLS AND COMPONENTS Headrest Assembly 1 Fixation light cable 2 Forehead rest band 3 Patient s eye height marker 4 Fixation light 5 Chinrest 6 Chinrest height a...

Страница 6: ...an electric insulated and fire resistant medical table top e g a refraction stand or combi unit Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any of the conten...

Страница 7: ...ETERS T TYPE AND R TYPES Tomey Applanation Tonometer T TYPE 1 Position the guide plate in the tonometer test bar support hole on the slit lamp 2 Lift Tonometer out of the packaging and assemble it by...

Страница 8: ...tact with the patient should be cleaned in accordance with these instructions prior to the examination Tomey recommends the use of disposable hygienic chinrest tissues on the Chinrest before the patie...

Страница 9: ...3 In the unlikely event of an LED failure please contact Tomey or your local distributor for guidance on the replacement procedure 7 2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT 1 Routinely clean...

Страница 10: ...ded for use in home environment Harmonic emissions IEC 61000 3 2 Class B Voltage fluctuations flicker emissions IEC 61000 3 3 Complies 9 2 ELECTROMAGNETIC IMMUNITY Guidance and manufacturer s declarat...

Страница 11: ...150kHz to 80 MHz field strengths should be less than 10 V m 9 3 RECOMMENDED SAFE DISTANCES Recommended separation distances between portable and mobile RF communications equipment and the Tomey Slit L...

Страница 12: ...al Electrical System as defined in EN IEC 60601 1 2006 The power supply forms a part of ME equipment Fuse Ratings and Quantity 2 5 amp anti surge Fuse current 2 5A Voltage rating V AC 250V Breaking ca...

Страница 13: ...k Patient Handle EP79 80070 11 PACKAGING AND DISPOSAL INFORMATION Applicable in the European Union and other European Countries with separate Collection Systems Disposal of old electrical and electron...

Страница 14: ...AA UK Freephone 0800 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 51500 Issue 5 Date of Issue 12 05 2021 SUPPLIER TOMEY GmbH Wiesbadener Stra e 21 90427 N rnberg Germany Tel 49 911 938 546 2...

Отзывы: