background image

 18

 19

EN

EN

SLIT LAMP TSL-7000Z by TOMEY

SLIT LAMP TSL-7000Z by TOMEY

Immunity 

test

IEC 60601
Test level

Compliance 

level

Electromagnetic environment 

– guidance

Portable and mobile RF communications equipment 

should be used no closer to any part of the Tomey 

Slit Lamp, including cables, than the recommended 

separation distances calculated from the equation 

applicable to the frequency of the transmitter.

Recommended  separation distance

Conducted 

RF IEC 

61000-4-6

6 Vrms 1 

50kHz to  

80MHz

6 V

d = 1.2 √ p

Radiated 

RF IEC 

61000-4-3

10 V/m 150kHz 

to 280MHz

10 V/m

d = 1.2 √ p 80MHz to 800 MHz 
d = 2.3 √ p 800MHz to 2.7GHz

Where p is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter 

manufacturer and d is the recommended separation 

distance in metres (m).
Field strengths from fixed RF transmitters, as 

determined by an electromagnetic site survey¹, 

should be less than the compliance level in each 

frequency range.²

Interference may occur in the vicinity of 

equipment marked with this symbol.

Note: At 80MHz and 800MHz, the higher frequency range applies. These guide lines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
1 Field strengths from fixed transmitters, such as base stations (cellular / cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with 

accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey 

should be considered. If the measured field strength in the location in which the Tomey Slit Lamp is used exceeds 

the applicable RF compliance level above, the Tomey Slit Lamp should be observed to verify normal operation. If 

abnormal performance is observed, additional measures may be necessary, such as re-orientating or relocating the 

Tomey Slit Lamp.
2 Over the frequency range 150kHz to 80 MHz, field strengths should be less than 10 V/m.

9.3  RECOMMENDED SAFE DISTANCES

Recommended separation distances between portable and mobile RF 

communications equipment and the Tomey Slit Lamp

The Tomey Slit Lamp is intended for the use in an electromagnetic environment in which 

radiated RF disturbances are controlled. The customer or the user of the Tomey Slit Lamp can 

help prevent electromagnetic interference by maintaining a minimum distance between mobile 

RF communications equipment (transmitters) and the Tomey Slit Lamp as recommended below, 

according to the maximum output power of the communications equipment.

Rated maximum output 

power of transmitter (W)

Separation distance according to frequency of 

transmitter (m)

150 kHz to 80MHz 

d = 1.2√ p

80MHz to 800MHz 

d = 1.2√ p

800MHz to 2.7GHz 

d = 2.3√ p

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended 

separation distance din metres (m) can be determined using the equation applicable to the 

frequency of the transmitter, where p is the maximum output power rating of the transmitter in 

watts (W) according to the transmitter manufacturer.

Note: At 80MHz and 800MHz, the separation distance for the higher frequency applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

9.4  TECHNICAL SPECIFICATIONS

Optical System

Binoculars type

Galilean 8° converging | Galilean 0° parallel

Magnification

Drum change x6, x10, x16, x25 & x40 | x10, x16 & x25

Eyepiece

x12.5

Field of view

34, 22, 14, 8.5 and 5.5 mm | 22, 14 & 8.5mm

PD range

8° converging eyepieces, 49 – 77mm
0° parallel eyepieces, 38 – 85mm

Objective lens focal distance

107mm

Objective lens convergence angle

13°

Slit Projection System & Base 

Slit Width

0-14mm continuously variable

Slit Length

14mm

Aperture diameters

0.2, 1mm square, 2, 3, 5, 9, 14 and reserved

Filters

Clear; red free; neutral density; blue;

Slit rotation

360° continuous, detents at 0° and 180°

Base travel

25mm Z-axis, 107mm X-axis, 110mm Y axis

Horizontal fine adjustment

12mm

Содержание TSL-7000Z

Страница 1: ...TSL 7000Z SLIT LAMP OPERATOR MANUAL Z Series...

Страница 2: ...IONS FOR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYSTEMS 1...

Страница 3: ...f the anterior eye segment The Tomey Slit Lamp is an instrument consisting of a light source that can be focused to shine a thin sheet slit of light into the eye It is used in conjunction with a biomi...

Страница 4: ...is disconnected LEDs can reach high temperatures in use allow to cool before handling Do not exceed maximum recommended exposure time Should the instrument suffer shocks for example should it accident...

Страница 5: ...f used cleaning materials 4 NAMES OF CONTROLS AND COMPONENTS Headrest Assembly 1 Fixation light cable 2 Forehead rest band 3 Patient s eye height marker 4 Fixation light 5 Chinrest 6 Chinrest height a...

Страница 6: ...an electric insulated and fire resistant medical table top e g a refraction stand or combi unit Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any of the conten...

Страница 7: ...ETERS T TYPE AND R TYPES Tomey Applanation Tonometer T TYPE 1 Position the guide plate in the tonometer test bar support hole on the slit lamp 2 Lift Tonometer out of the packaging and assemble it by...

Страница 8: ...tact with the patient should be cleaned in accordance with these instructions prior to the examination Tomey recommends the use of disposable hygienic chinrest tissues on the Chinrest before the patie...

Страница 9: ...3 In the unlikely event of an LED failure please contact Tomey or your local distributor for guidance on the replacement procedure 7 2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT 1 Routinely clean...

Страница 10: ...ded for use in home environment Harmonic emissions IEC 61000 3 2 Class B Voltage fluctuations flicker emissions IEC 61000 3 3 Complies 9 2 ELECTROMAGNETIC IMMUNITY Guidance and manufacturer s declarat...

Страница 11: ...150kHz to 80 MHz field strengths should be less than 10 V m 9 3 RECOMMENDED SAFE DISTANCES Recommended separation distances between portable and mobile RF communications equipment and the Tomey Slit L...

Страница 12: ...al Electrical System as defined in EN IEC 60601 1 2006 The power supply forms a part of ME equipment Fuse Ratings and Quantity 2 5 amp anti surge Fuse current 2 5A Voltage rating V AC 250V Breaking ca...