Calibration Procedure
(1) Precautions for use
1) Be careful not to drop weights on your foot.
2) Place a weight gently.
3) Start calibration at the minimum value.
4) Before calibration, secure the main unit in fixed position to prevent it from moving during
calibration.
5) Hook wire on the calibration lever securely.
(2) Calibration procedure
1) Place TDT3-G on the measurement table in a horizontal position.
2) Turn on the power of TDT3-G. (After the power is turned on, leave the unit as it is for 30
minutes or more for stabilization.)
3) Set TDT3-G in RUN mode.
4) Insert the square drive of the calibration lever into the TDT3-G sensor portion so that the
calibration lever is leveled.
5) Hook the wire on the calibration lever, and press "C" key on TDT3-G to make zero
adjustment.
6) Refer to the table on next page (Weights Information for TDT3-G Calibration), and check
the weight. Place weight on the TDT3-G accordingly, and proceed calibration.
10
Calibration
roller
TDT3-G
main unit
Calibration
wire
Position of calibration
roller for TDT60CN3-G
Position of calibration
roller for TDT600CN3-G
Calibration lever
for TDT60CN3-G
Calibration lever
for TDT600CN3-G