![Tohnichi TDT3-G Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/tohnichi/tdt3-g/tdt3-g_operating-instruction_1124791022.webp)
21
n
D
IG
IT
A
L
T
O
R
Q
U
E
S
C
R
E
W
D
R
IV
E
R
T
E
S
T
E
R
M
O
D
E
L
T
D
T
3
-G
8-1. Printer Output
Connect TDT3-G and the printer (EPP16M2) using the cable (No. 382). Set the communication
output format to “Prn” and other communication settings compliant with printer settings (Refer to
chapter 6 for details.)
* Do not use the cable for printer and the USB cable at the same time.
• Communication settings of EPP16M2
Data format
: Compliant with RS232C
Data transmission
: Start-top synchronization
Baud rate
: 2400bps
Data length
: 7bit
Stop bit
: 1bit
Parity
: None
(1) Printing one by one
After measurement, press Memory/Reset key to print the measured data. In this mode, the
results of statistical processing are not printed. In Auto Memory/Reset mode, printing starts
automatically.
(2) Printing a certain range of data
Using
s
or
t
key, select the last count value in the print range and press STAT key, then “ST”
will be displayed. Using
t
or
t
key, select the first count value in the print range and press
STAT key. Then, press
s
key while “n” is displayed. The measured value, sampling quantity
“n”, maximum value “MAX”, minimum value “MIN” and mean value “x” of the selected range
will be printed out.
(3) Printing only statistical processing value
Using
s
or
t
key, display the last count value in the print range and press STAT key. Then,
“ST” will be displayed. Using
s
or
t
key, display the first count value in the print range and
press STAT key. Press
t
key while “n” is displayed. Only the sampling quantity “n”, maximum
value “MAX, minimum value “MIN” and average value “x” of the selected range will be printed
out.
8
External Output