Tohnichi TDT3-G Скачать руководство пользователя страница 19

7-1-1. Operating Temperature

Measuring  should  be  conducted  at  temperature  in  18-28°C  range.  Temperature  fluctuation

should be within ±1°C.

7-1-2. Check the Following before Calibrating the Torque Screwdriver.

(1) Torque screwdriver tester should be set at stable working table.

(2) For  calibration  of  direct  reading  torque  screwdriver,  make  sure  to  read  the  value  from

upright position to avoid reading errors by angle.

(3) For  calibration  of  click-type  torque  screwdriver  (RTD/LTD),  apply  training  torque  at  the

maximum value for 5 times without measuring prior to the official calibration. Do the same

training for the other direction as well.

(4) For calibration of direct reading torque screwdriver (FTD, STC), apply training torque to the

max  torque  value  once  and  release  the  torque.  Then  set  the  digital  display  back  to  zero.

Repeat the same procedure for the other direction as well.

7-2. Attention for Measuring Torque Screwdrivers

(1) For  measuring  click-type  torque  screwdrivers  (RTD/LTD),  apply  torque  slowly  and  evenly.

Except  for  preset  type  (such  as  RNTD/NTD),  click-type  torque  screwdrivers  (such  as

RTD/LTD) must measure from the lower point, and adjust to respective torque value.

(2) For  measuring  direct-reading  type  torque  screwdriver  (FTD/STC,  etc),  use  torque

screwdriver tester to apply torque to the measurement value. If the applied torque exceed

the measurement point, it needs to be redone. 

* LTA loading device is an optional kit.

18

7

Measuring a Torque Screwdriver (Recommended)

n

D

IG

IT

A

L

 T

O

R

Q

U

E

 S

C

R

E

W
D

R

IV

E

R

 T

E

S

T

E

R

   

M

O

D

E

L

 

T

D

T

3

-G

Содержание TDT3-G

Страница 1: ...y and safely please read this manual carefully before use If you have any question about the product and its operations please contact your nearest distributor or TOHNICHI MFG CO LTD TDT3 G TDT3 G Mod...

Страница 2: ...le gas is present It may cause electric shock or fire 11 Be sure to use the specified accessories or options Do not use any accessory or option other than those specified in this operating instruction...

Страница 3: ...ay cause accident or damage 6 Check for any damaged parts Before use check the tester accessories and other parts for damage and make sure that it operates normally and fulfills the specified function...

Страница 4: ...Peak Hold Mode 10 5 3 Selecting Upper and Lower Limit Value 10 5 4 OK NG Judgment Data Memory 10 5 5 Auto Memory Reset 10 5 6 Deletion of Data 11 5 7 Statistical Processing Function 11 5 8 Auto Zero A...

Страница 5: ...f samples maximum value minimum value or mean value of the stored data can be calculated and displayed Judgment function allows users to register upper lower limit torque value and have TDT3 G give ju...

Страница 6: ...amp block Fixing knob Clamp tool Standard bit bit S AC adapter BA 6 Power connector conversion adapter C type Operating instruction Operation sticker When purchasing TDT3 G package includes STA loadin...

Страница 7: ...per Lower Limit Value Setting 10 kinds Zero Adjustment Auto Reset Manual Auto 0 1 5 0 sec selectable Data Output RS232C compliant USB B type serial communication Square Drive Hex 6 35 male minus 0 7mm...

Страница 8: ...it bottom to upper side of clamp height 80 Using bit S From bit bottom to upper side of clamp height 105 Using standard bit MEMORY RESET SWITCH Applicable grip dia 7 to 50 Clamp tool to put Standard b...

Страница 9: ...er limit value 3 Set A J display Lower limit value It shows the limit value registration Set A J It also shows Lower limit value 4 t mark display RUN RUN mode is selected PEAK PEAK mode is selected M...

Страница 10: ...ection 10 MEMORY RESET switch After giving judgement to the measured data it moves the counter forward by one 4 2 Power Source and Output USB terminal RS232C terminal Power jack Power switch 1 Power j...

Страница 11: ...suring condition with previous upper lower limit value settings For setting upper lower limit value refer to chapter 6 Various Settings 5 4 OK NG Judgment Data Memory As MEMORY RESET button is pushed...

Страница 12: ...vice 5 7 Statistical Processing Function Number of sample Max Minimum Mean value 1 Using st key select the last counter of the data range you want to calculate 2 Push STAT key then Stt appears on the...

Страница 13: ...Err1 5 appears Turn off the power and turn it on without touching any keys If Err disappears it is restored If Err continues to stay contact your nearest distributor or TOHNICHI to seek assistance whe...

Страница 14: ...000 60 000 000 00 600 00 6 Lower limit value B Lo SEt b 00 000 000 00 00 000 60 000 000 00 600 00 23 Upper limit value J HI SEt J 00 000 000 00 00 000 60 000 000 00 600 00 24 Lower limit value J Lo SE...

Страница 15: ...out setting Selecting measuring unit default cN m Select measuring unit Use st key to select unit and push STAT key to decide If you press C key it returns to measuring display without setting Registe...

Страница 16: ...measuring display If you want to cancel settings push C key to return to the measuring display without setting 6 3 Parameter Settings Setting mode At no load condition push C STAT MD key then it turns...

Страница 17: ...ut and PC for PC output Push STAT key to decide It proceeds to baud rate setting If you press C key it returns to measuring display without setting 2 Baud rate setting default 2400bps Set baud rate Us...

Страница 18: ...ax 4 digit Using st key to select 1000 or 99 and push STAT to confirm If you press C key it returns to measuring display without setting Push STAT key to switch memory mode clear the memory data and p...

Страница 19: ...ct reading torque screwdriver FTD STC apply training torque to the max torque value once and release the torque Then set the digital display back to zero Repeat the same procedure for the other direct...

Страница 20: ...to zero adjustment Remove the torque screwdriver from the main body and push C to activate auto zero adjustment 7 Set the bit according to the measuring model See below table to see which bit suit you...

Страница 21: ...11 process to measure Table STA loading device and bit selction 1 Put the torque screw driver through the loading device Torque screwdriver Loading center black Hold the external rim black of the loa...

Страница 22: ...Auto Memory Reset mode printing starts automatically 2 Printing a certain range of data Using s or t key select the last count value in the print range and press STAT key then ST will be displayed Us...

Страница 23: ...1000 In case of memory mode Memory counter judgement result upper lower Limit value symbol torque unit for no judgement Torque value Number of sample Max value Min value Mean value Torque value unit...

Страница 24: ...top bit 1bit Parity None Even Odd 1 Printing one by one Push Memory Reset switch or automatically print by Auto memory Reset function 2 Printing all at one time Set the last counter of the data range...

Страница 25: ...A3 Error list 9 Error Message Meaning Measures to Take Err1 5 Switch is kept being Turn off the power Turn it on without touching any key If the error is pushed cancelled it can be used If it still sh...

Страница 26: ...is turned on leave the unit as it is for 30 minutes or more for stabilization 3 Set TDT3 G in RUN mode 4 Insert the square drive of the calibration lever into the TDT3 G sensor portion so that the cal...

Страница 27: ...G Measuring Loading Measuring Loading Point Weights Remarks Point Weights Remarks kgf cm kg kgf cm kg 0 2 0 1 Scale Pan 2 0 2 Scale Pan 100g 2 1 0 Scale Holder 20 2 0 Scale Holder 1kg 3 1 5 Scale Hold...

Страница 28: ...f in kg 3 0 1080 Scale Pan 8 0g 30 0 2160 Scale Pan 116 0g 20 0 7200 Scale Pan 620 0g 200 1 440 Scale Holder 440g 40 1 440 Scale Holder 440g 400 2 880 Scale Holder 1 880kg 60 2 160 Scale Holder 1 160k...

Отзывы: