(4) Displaying the measured data
Press key to reverse the memory counter to check the previous measured data.
When the memory counter is 001, press key one more time to set it 000, then it
becomes the RUN mode.
When the memory counter is 000, press key one more time, then the memory
counter goes to 999 and the measured data will be shown accordingly.
(5) Outputting the measured data
●
Data output (1 reading)
Use the dedicated communication cable to connect CTA2-G to PC or printer.
Use key to show the data to output and press MEM key to transfer.
●
Data output (a set range of data at one time)
Use the dedicated communication cable to connect CTA2-G to PC or printer and set
the memory counter to the upper limit of the output data range to transfer.
Press MD key to confirm the upper limit, then the display proceed to the lower limit
setting.
Use key to set the lower limit of the output data range. Press MD to confirm
the lower limit and the display shows the number data to transfer (Press C key to
cancel).