Model: TTT-710
4.5 Mounting the electrode cartridge, electrode, and titration nozzle
- 32 -
Types of Electrode Cartridges Table
Name
(Number of mountable
electrodes and nozzles)
*
1
Model
Number of
usable table
specimens
Usable electrode type
*
2
Appearance
1CH cartridge 1(X)
(Electrode:1 Nozzle: 1)
*
3
7505010K All number of
specimens
X
R
J (white adapter use)
1CH cartridge 2(X)
(Electrode:1 Nozzle: 2)
7505020K
12 specimens
18 specimens
36 specimens
X
R
J (white adapter use)
1CH cartridge 3(X)
(Electrode:1)
7506840K
All number of
specimens
X
R
J (White adapter use)
2CH cartridge 1
(Electrode:2 Nozzle: 1)
6597970K
12 specimens
18 specimens
36 specimens
J
ION (silver adapter
use)
2CH cartridge 2(X)
(Electrode:2)
7505030K
12 specimens
18 specimens
36 specimens
X
R
J (white adapter use)
ION (silver adapter
use)
2CH cartridge 3
(Electrode:2 Nozzle: 2)
6597940K
12 specimens
18 specimens
X
R
J (white adapter use)
ION (silver adapter
use)
Multi cartridge
(Electrode: 4 Nozzle: 4)
6597980K 12
specimens
J (max. 2)
ION (2)
(Up to 4 can be
mounted by using the
silver adapter)
Sample suction cartridge
(Nozzle: 1)
6597990K All number of
specimens
−
Nozzle
*1: A temperature sensor can be installed by using the accessory blue adapter at the nozzle mounting hole.
However, with 1CH cartridge 1(X), it can be used only when the 12, 18, or 36 specimens tables are used.
*2: Electrode types “X, R, J, or ION” described in the table indicate the usable electrode type classification. For details,
refer to the Usable Electrode Type Classification Table.
*3: When 60 or 100 specimens table is used, only the micro titration nozzle (AST-P008) can be mounted.