Model: TTT-710
4.5 Mounting the electrode cartridge, electrode, and titration nozzle
- 31 -
4.5 Mounting the electrode cartridge, electrode, and titration
nozzle
The electrode cartridge is different depending on the number of table plate specimens, kind of
electrode used, and the number of titration nozzles used. Use the electrode corresponding to the
application.
>>
“Types of Electrode Cartridge Table”
Mounting the Electrode Cartridge, Electrode, and Titration Nozzle
①
Mount the electrode cartridge.
⋅⋅⋅⋅⋅⋅
After loosening the knurled knob at the end of the
electrode arm of the product main body, hold the electrode cartridge with the dent facing upward
and insert the cartridge into the end of the electrode arm and fix it by tightening the knurled knob.
②
Mount the electrode and titration nozzle.
⋅⋅⋅⋅⋅⋅
Loosen the electrode fixing bolt of the
electrode cartridge, insert the electrode and titration nozzle to be used, and fix the electrode by
tightening the electrode fixing bolt. (The electrode cartridge accessory adapter may be necessary
depending on the electrode to be used. Refer to the “Types of Electrode Cartridge table” and
“Usable Electrode Type Classification Table” for the adaptors corresponding to the electrodes
used.)
③
Clamp the cable, etc.
⋅⋅⋅⋅⋅⋅
Clamp the electrode cable and the tube from the titration nozzle with
the clamp at the top of the electrode arm.