![Tissue-Tek TEC 5100 Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/tissue-tek/tec-5100/tec-5100_operating-manual_1119157007.webp)
Revised 07/12/11
1.3
SPECIFICATIONS
EMBEDDING
CRYO
MODULE
MODULE
Ambient Operating
10° – 40° C
10° – 40° C
Temperature and
Relative Humidity Range
30% – 85% RH,
30% – 85% RH
(noncondensing)
(noncondensing)
Dimensions/Weight
Width
57.5 cm (22.6 in)
Width
33.0 cm (13.0 in)
Depth
64.2 cm (25.3 in)
Depth
61.7 cm (24.3 in)
Height
37.7 cm (14.8 in)
Height
37.7 cm (14.8 in)
Weight approx. 26 kgs (57 lbs) Weight approx. 22 kgs (48 lbs)
Power Required
Model No. 5238:
100 VAC ±10% or less,
100 VAC ±10% or less,
@ 2.0 amps 50/60 Hz
@ 9.0 amps 50/60 Hz
Model Nos. 5100, 5235:
115 VAC ±10% or less,
115 VAC ±10% or less,
@ 2.0 amps 60 Hz
@ 9.0 amps 60 Hz
Model Nos. 5229, 5232:
220 – 240 VAC ±10% or
220 – 240 VAC ±10% or
less, @ 2.0 amps 50 Hz
less, @ 9.0 amps 50 Hz
Chamber Capacity
Paraffin Chamber:
approx. 4 liters
(no chambers)
Left and Right Chambers:
approx. 1.5 liters
Temperature Range
Paraffin Chamber, Right and
Cooling Plate:
Available
Left Chambers, Hot Plate:
–10°C to 0°C
50°C – 75°C
Electric Forceps(optional):1 to 5
Factory Set Temperature
Paraffin Chamber, Left and
Cooling Plate:
Right Chambers, Hot Plate:
– 5°C
62°C
Cold Spot: approx. 15°C at
25°C room temp. and
40% R.H.
Electric Forceps(optional): 3
Circuit Breaker/ON-OFF
15 amps
Switch
Memory Backup
Long-life battery
Safety Standards
Tested and listed by Intertek Testing Services (ITS)
Complies with: UL 3101-1, 1st ed., IEC 1010-1 and
CAN/CSA-C22.2 No. 1010.1-92