DAILY MAINTENANCE
Embedding Module
REPLENISH PARAFFIN
WARNING: MOLTEN PARAFFIN MAY
CAUSE BURNS. DO NOT OVERFILL
PARAFFIN CONTAINER.
Q
2 parts dry paraffin = 1 part molten paraffin.
Q
The paraffin chamber holds 4 liters (1.06 gal)
of molten paraffin.
Q
The paraffin bath (either transfer tray or
chamber, whichever you are using) should be
half-filled or about .9 liter (.24 gal) with
paraffin.
NOTE:
Refer to Section 3 for complete instructions
regarding “Adding Paraffin.”
REMOVE SPILLED PARAFFIN/DEBRIS
WARNING: XYLENE IS FLAMMABLE.
Q
Using the provided plastic scraper, remove
solidified paraffin from the gray insulated
work space and hot plate area.
Q
As necessary, wipe the insulated work space
and hot plate area with a xylene-dampened
cloth.
Q
Clean the grooves surrounding the hot plate
areas with a xylene-dampened cotton swab.
Cryo Module
Turn off compressor (press CRYO key on
Embedding Module Control Panel).
Allow cooling plate to reach ambient temperature.
Remove condensate from cooling plate (use an
absorbent cloth/towel).
DISINFECTION
After performing daily maintenance, you may disin-
fect the surfaces and keypads (if desired) as follows:
Q
Wipe area using a soft cloth dampened in alcohol
(70 – 75% solution isopropyl or ethyl).
Q
Do not wipe dry. Allow area to air dry to maxi-
mize disinfection.
7.2
Revised 07/12/11
CARE OF THE INSTRUMENT