EMBEDDING
MODULE ERROR
CONDITION
CAUSE
SOLUTION
Power indicator will not
Power outage occurred
Check to make sure:
illuminate
or Cryo Module has
Q
power cord is securely plugged into a “live”
malfunctioned.
Q
power outlet.
Q
the power switch on the front panel of Cryo
Module is set to ON.
Q
The cable connecting both modules is firmly
inserted at both ends.
If the problem persists, contact Customer Support.
Alarm (tone) sounds for
Temperature is
Press “SET” key. Highlight menu item
10 seconds;
out-of-range by more
“6. ERROR LIST” and press ENTER. Record
than 5°C (9°F)
the error indication and press EXIT.
If more than one error code is shown, use the
arrow keys to move the cursor to highlight the next
error code number, and record the error indication.
Embedding Module
Check to make sure:
status never reaches
Q
if you have recently added cold paraffin, base
<READY>
Q
molds, install a cool forceps warmer, etc.,
Q
that may have caused more than a 5°C drop
Q
in an area temperature. If so, wait a few minutes
Q
for the temperature to stabilize and the error
Q
condition will then cease.
Q
that at least 76 mm (3 inches) of space exists
Q
between the wall or other surface and the
Q
back panel of the module.
Q
that you have checked the temperature of the
Q
room. Adjust the temperature if necessary
Q
the AUTO/MAN key is set to MANUAL and
the HEAT key is pressed ON.
Q
The AUTO/MAN key is set to AUTO and the
ON and OFF times for the current day are set
properly.
If the error condition persists, contact Customer
Support.
Revised 07/12/11