![Tiger Optics HALO M7000 Series Скачать руководство пользователя страница 49](http://html.mh-extra.com/html/tiger-optics/halo-m7000-series/halo-m7000-series_instruction-manual_1115521049.webp)
HALO Users Manual
Revision G
41
TAUSTD
Purpose
Read the
standard deviation
in the on-peak ringdown time
Syntax
“TAUSTD”
Arguments
none
Remarks
The
standard deviation
in the on-peak ringdown time is calculated by
the HALO while in the MEASURE mode.
Range 0.01 to 10
Example
Command : “TAUSTD”
Reply:
“0.098”
Errors
!BAD VALUE
The value
f
is invalid
TAUEMPTY
Purpose
Read the off-peak ringdown time.
Syntax
“TAUEMPTY”
“TAUEMPTY”
Remarks
The off peak ringdown time is normally calculated by the HALO’s
TUNE mode.
Example
Command : “TAUEMPTY”
Reply:
“72.123”
Errors
none
TAUESTD
Purpose
Read the standard deviation in the off-peak ringdown time
Syntax
“TAUESTD”
Arguments
f
: the standard deviation of the ringdown time in microseconds
Remarks
The standard deviation in the off peak ringdown time is
calculated by the HALO’s TUNE mode.
Example
Command :
“TAUESTD”
Reply:
“0.098”
Errors
!BAD RANGE
The value
f
is invalid