Gentier Real-Time PCR System
Click
< OK >
and the application software will automatically divide the replicate group for all
the selected samples;
User can click
< Cancel >
and the application software will give up the current replicate
setting and automatically return to the sample setting interface.
▶
Sample ID
:
User can select a well in the sample setting area and set the sample ID for the selected well in
the
Sample ID
input box; click
< Apply >
to confirm the sample ID;
User can also select multiple wells in the sample setting area, and let the application software
to set the corresponding sample IDs. Please enter the starting ID number in the
Sample ID
input box and click
< Apply >
; the choose direction dialog box will pop-up, user could set the
Horizontal
or
Vertical
direction and then click
< OK >
, the application software will
automatically set the sample IDs for the selected wells according to the direction setting.
▶
Unique ID
:
User can select a well in the sample setting area and set the unique ID for the selected well in
the
Unique ID
input box; click
< Apply >
to confirm the unique ID;
User can also select multiple wells in the sample setting area, and let the application software
to set the corresponding unique IDs. Please enter the starting ID number in the
Unique ID
input box and click
< Apply >
; the choose direction dialog box will pop-up, user could set the
Horizontal
or
Vertical
direction and then click
< OK >
, the application software will
automatically set the unique ID for the selected wells according to the direction setting.
Warning
: The
Sample ID
and
Unique ID
should be at least double-digit, and last two digits
must be numbers.
Reminding
: The
Sample ID
setting is primarily intended to facilitate operators to identify
and differentiate samples.
Samples can only be identified and confirmed according to the
Unique ID
.
▶
Reference Dye
:
User can set the reference dye from the
Reference Dye
drop-down list.
▶
QC Failure Condition Setting
: after completing the sample well setting, please click
icon and the QC failure condition setting dialog box will pop up
automatically, as shown in figure D-6.7.