Gentier Real-Time PCR System
Initial Time
: the initial value of the elongation time change range.
Target Time
: the target value of the elongation time change range.
Delta Time
: the temperature change(°C) per cycle.
Start Cycle
: the cycle number after which the time change is started.
Reminding
: the initial and target time setting range is 1s~60min.
Reminding
: the delta time setting range is 1s~10min.
Reminding
: the star cycle range is 1~ max cycle number of the current stage.
Reminding
: the other parameter settings are same with the standard step mode.
4)
If the
Gradient
step mode option is selected, the step edit window will be shown as figure E-
3.6.
Figure E-3.6 Temperature Setting sub-tab - Step Edit window - Gradient step mode
▶
The gradient step mode allows the Gentier sample to adopt different temperatures. After
setting the temperature center value and the temperature offset value for the current
gradient step, the system will automatically calculate the gradient temperatures.
▶
User can press the input boxes of
Temp. Center
and
Temp. Offset
to edit the corresponding
parameters for the current gradient step. User could also and press
< Details >
to view the
specific gradient temperatures of each row, as shown in figure E-3.6a.