5.2 Survey of Commands
By the serial interface of the VD10 the following functions are available:
Command
Code
Data Type
Function
Type
T
STRING
read
Measurement Value
M
FLOAT
read
Control start/stop
A, a
BOOLEAN
read A, write a
Keylock
K, k
BOOLEAN
read K, write k
For reading transducer type or measurements of channel 1 to 4 the
corresponding address settings "001" to "004" have to be used.
The following data strings are sent as possible answer to type queries:
"V10205"
measuring channel VD10 with VSR
"V10206"
measuring channel VD10 with VSP
"V10207"
measuring channel VD10 with VSM
"V10208"
measuring channel VD10 with VSH
For commands "a/A" and "k/K" address "001" can be used.
Examples:
Action
Telegram to
transducer
Answer telegram
from transducer
Read pressure measurement
Of channel 4
"004Ma
C
R
"
"004M260014N
C
R
"
(
2.6x10-6 mbar)
Start Control
"001a1c
C
R
"
"001a1c
C
R
"
For further information please refer to the separate description of the Thyracont
communication protocol.