Benchtop Serial I/O
7/30/2015 3:30 PM
Page 9
Change RH @Pc Setpoint
R1=
Description
Changes the RH @Pc setpoint to a given value.
Prerequisites
Requires Benchtop firmware revision 1.05 or greater.
Syntax
R1=
setpoint
Parameters
setpoint
The
setpoint
value is the %RH that the system should generate.
Remarks
Control Mode Changed
Sending
R1=
setpoint
also changes the control mode to %RH @Pc
(mode 1). In this mode, the %RH @PcTc and Saturation Pressure
setpoints vary as system temperatures and pressures change.
Data Returned
None. The only response is a carriage return/linefeed terminator.
Examples
The
R1=
setpoint
command is used to set the %RH @Pc to 90%RH.
computer sends
R1=90
<CR>
benchtop responds
<CR><LF>
The
<CR><LF>
response signifies completion of the command.