3-10
An incorrect code prevents access and returns to step 3. A correct code results in the
display of calibration coefficients.
4. Using the [NEXT] and [PREV] keys, view any or all of the remaining coefficient and
parameter screens.
5. To edit a particular displayed value, press [EDIT]. The cursor begins flashing at the left
of the first parameter displayed.
Using [±±], ['E'], arrow keys, and the numeric keys as necessary, change any or all
displayed values as desired. Then press <ENTER>.
6. After Editing or Viewing, press [DONE]. Then from the next menu, press [DONE]
again. The system reinitializes back to the Control/Display screen.
SATURATION TEMP, Ts
Zero
[-.0851657
]
Span
[ .0200085
]
Lin
[ 4.85408E-09 ]
Avg
[ 10
]
Last Cal Date
08/10/12
EDIT
NEXT
PREV
DONE
MASS FLOW RATE
Zero
[ 0
]
Span
[ 1.6E-3
]
Lin
[ 0
]
Avg
[ 10
]
Last Cal Date
08/10/12
EDIT
NEXT
PREV
DONE
MASS FLOW RATE
Zero
[~0
]
Span
[ 1.6E-3
]
Lin
[ 0
]
Avg
[ 10
]
Last Cal Date
08/10/12
+ +
- -