Benchtop Serial I/O
7/30/2015 3:30 PM
Page 12
Change Saturation Temp Setpoint
TS=
Description
Changes the Saturation Temperature setpoint to a given value.
Prerequisites
Requires Benchtop firmware revision 1.05 or greater.
Syntax
TS=
setpoint
Parameters
setpoint
The
setpoint
value is the Saturation Temperature, in °C
only
, that the
system should control at.
Remarks
Precautions
Chamber temperature follows Saturation Temperature. When
increasing Saturation Temperature, always do so at low humidity so as
to prevent condensation on cool devices within the chamber. Allow
time for temperatures of the devices to rise before raising humidity.
Data Returned
None. The only response is a carriage return/linefeed terminator.
Examples
The
TS=
setpoint
command is used to set the Saturation Temperature to
10°C.
computer sends
TS=10
<CR>
benchtop responds
<CR><LF>
The
<CR><LF>
response signifies completion of the command.