3-8
3.2.8.1 Maintenance Power-Off
Before maintenance of any kind, other than calibration, the following must first be
performed:
A) Press [STOP], if system is currently RUNNING, to close all
valves and disable control.
B) Press main rear panel Power switch to OFF.
C) Disconnect console from primary power.
D) Shut off and/or disconnect air supply.
3.2.9 Filling the Reservoir
The reservoir level is graphically displayed on the control/display screen by the computer
control system. When the level displayed is empty or prior to a long term test the reservoir
should be filled. The 2500 humidity generator holds approximately one gallon of distilled
water which can last from 12 hours to over 9 months of continuous use, depending on the
temperature and humidity at which it is operated (reference section 1.6.5)
Note:
Reservior/presaturator system has been
EMPTIED
by factory
prior to shipment.
To fill the reservoir:
A) Press [STOP], if the generator is running, and allow all pressure to be vented.
B) Remove the reservoir fill port cap located on the top left panel of the system.
C) Add up to 1 gallon (3.8 liters) of double distilled water until the indicator on the
front panel reads full. Use the reservoir fill funnel or a small funnel may be
useful. Add the distilled water slowly so as not to overfill the reservoir.
D) Replace fill port cap, finger tighten, and proceed with operation.