Wavefront Select Switch
The switch below the graph determines the wavefront type that is displayed.
'
Measured
' shows the wavefront which is directly calculated from the measured
spot deviations using a 2-dimensional integration.
'
Reconstructed
' will display a wavefront that is reconstructed using all or only a
selected set of the determined Zernike coefficients. The advantages of this
display is:
·
The wavefront surface is smoothed because noise is cancelled out when only a
few low order Zernike modes are selected for reconstruction.
·
Lowest order Zernike terms which are always present but are of less interest
(for instance piston, tip, tilt) can be cancelled in order to see only the higher
order modes of interest.
·
Only particular Zernike modes that are enabled within the
panel can be displayed and analyzed separately.
'
Difference
' displays the difference between the measured and reconstructed
wavefront and is therefore an indicator of the fit error.
Note:
The actual switch setting is also important to the calculated wavefront
statistics
PV, RMS
and
wRMS
) because the
displayed values refer to the selected wavefront (Measured, Reconstructed or
Difference).
Direct rotation, zoom and shifting of the 3D graph
You can zoom and shift the displayed 3D beam profile using the mouse and
modifier key Alt and Shift:
To
rotate
the graph about both longitude and latitude angles just click and hold
the
left mouse button
on the graph and drag it.
To
zoom
press and hold
Alt
key, then click and hold the
left mouse button
while
moving the mouse.
Shifting
the graph within the drawing area is done with pressing the
Shift
key
and holding the
left mouse button
while dragging the mouse.
Double click
on the graph to see the '
Wavefront Graph Configuration
' panel.
Содержание WFS150
Страница 1: ...2007 Operation Manual Thorlabs Instrumentation Optical Wavefront Sensor WFS150 WFS150C...
Страница 2: ...Version Date 1 2 December 2007 2007 Thorlabs GmbH 2007 Thorlabs GmbH...
Страница 6: ...General WFS150 WFS150C Manual Part I...
Страница 10: ...Getting Started WFS150 WFS150C Manual Part II...
Страница 25: ...Description WFS150 WFS150C Manual Part III...
Страница 35: ...Operating Instruction WFS150 WFS150C Manual Part IV...
Страница 103: ...Service and Maintenance WFS150 WFS150C Manual Part V...
Страница 107: ...Listings WFS150 WFS150C Manual Part VI...
Страница 110: ...Appendix WFS150 WFS150C Manual Part VII...
Страница 116: ...114 WFS150 WFS150C Manual 2007 Thorlabs GmbH Typical Response of the CCD camera...
Страница 117: ...115 Appendix 2007 Thorlabs GmbH 7 4 Wavefront Sensor Drawing...
Страница 123: ......