© 2017 Thorlabs GmbH
5 Operating Instruction
17
Cascading the interlock of multiple TXP5016 mainframes
Up to five TXP5016 mainframes can be cascaded in order to use a
common interlock.
The interlock opens and the safety function of all cascaded
mainframes is triggered as soon as the current is interrupted. In
other words, an active-open switch should be placed then between
contacts 5 and 9 of the first mainframe.
Note
When cascading two or more TXP5016 interlocks, only one constant
current source must be used!
Содержание TXP5016
Страница 1: ...Thorlabs Instrumentation TXP5016 Operation Manual 2017 ...
Страница 2: ...Version Date Item No 3 1 5 31 Jan 2017 M0009 510 102 Copyright 2017 Thorlabs GmbH ...
Страница 9: ... 2017 Thorlabs GmbH 2 Getting Started 7 Allow the installation Then follow the the screenshots below ...
Страница 10: ... 2017 Thorlabs GmbH 8 TXP5016 ...
Страница 42: ......
Страница 43: ......
Страница 44: ...www thorlabs com ...