41
T-Cube Position Aligner
13) It is advisable to click the ‘Persist Settings To Hardware’ box. These settings will
then be loaded on each power up cycle.
Fig. 5.6 Piezo Driver Settings - Rev 2 Hardware
14) Click ‘OK to save the settings.
15) On the GUI panel, turn the ‘Output’ control to give a reading of 0V on the display.
16) Click the ‘Settings’ button on GUI of the Piezo Driver fitted in bay 2. The Settings
panel is displayed
17) Make the following parameter settings, as shown in Fig. 5.6
Output Voltage Range -
For the ASM003 mirror shown in Fig. 5.5
,
set
75V
, (for
other piezo mirrors, set 75V, 100V or 150V whichever is applicable)
.
Drive Input Source (Open Loop) -
Select
Vin + SW + Pot
Analogue Input Source (EXT IN) -
Select
Hub Channel 1
Feedback Loop Mode -
Select
Open Loop
18) It is advisable to click the ‘Persist Settings To Hardware’ box. These settings will
then be loaded on each power up cycle
19) On the position aligner GUI, click the ‘Settings’ button to display the Settings
panel.