3.2 commands
Temperature module TED8xxx / page 47
":CALTB:MAX?"
Reading the maximum B
val
allowed
[:CALTB:MAX
<NR3><LF>]
":CALTR:MAX?"
Reading the maximum R
0
allowed
[:CALTR:MAX
<NR3><LF>]
":CALTT:MAX?"
Reading the maximum T
0
allowed
[:CALTT:MAX
<NR3><LF>]
#
Refer to section 2.2.6, “Calibrating the thermistor” on page 34)
NOTE
The selection on how the sensor calibration is done (exponential method
or Steinhart-Hart method) is done by the order in which you transmit the
coefficients.
If the last transmitted calibration command belongs to the exponential
method (see above), then the calculation is also done with the exponential
method.
If the last command was a Steinhart-Hart parameter, then this method is
chosen.