2.2 Functions in the channel menu
Temperature module TED8xxx / page 34
2.2.6 Calibrating the thermistor
2.2.6.1 Select the calculation method
If the relation between thermistor resistance and corresponding thermistor
temperature is known all inputs and outputs can be calibrated in °C.
Two methods to calculate the resistance from temperature are implemented:
•
The exponential method
•
The Steinhart-Hart method
In the channel menu you can choose between those two methods and enter the
corresponding parameters:
Select the line
Steinh.-Hart
resp.
Exponential
The right softkey toggles between the two methods.
Select the desired type and press
to make setting valid.