3.1 General notes on remote control
Temperature module TED8xxx / page 43
3 Communicating with a control computer
3.1 General notes on remote control
The description of the mainframe of the PRO8000 (-4) / PRO800 includes all
instructions of how to prepare and execute the programming of the system via
IEEE 488 computer interface.
Special operation features of a TED8xxx temperature module are described here.
#
(Refer to chapter 2, "Operating the TED8xxx" starting on page 26)
NOTE
All analog values are read and written in SI units, i.e. A (not mA), W (not
mW) etc. Letters may be written in small or capital letters.
!
Attention
!
Before programming a temperature module first set the limit value of
the TEC current I
LIM
(hardware limit) for the used TEC element with a
screwdriver.
The corresponding potentiometer is marked I
LIM
and is situated at the
front panel of the TED8xxx module.
The value I
LIM
is continuously measured by the PRO8000 (-4) /
PRO800 and can be checked in the sub-menu of the TED8xxx during
setting.
#
(Refer to chapter 1.6.1, "Setting the limit values" starting on page 13)